Oil Sampling Procedures
STANDARD OIL SAMPLING SOP WALKTHROUGH
This video demonstrates pulling an oil sample from our low pressure sample probe located on USPI Filtration systems.
NH3 Compressor OIL SAMPLING SOP WALKTHROUGH
This video demonstrates pulling an oil sample from our NH3 rated sample probe located on USPI Filtration systems.
Please select an option below:
These pages contain PDF files with step-by-step instructions for proper oil sampling technique. Each page gives instructions for a particular type of sample valve or equipment. Click on a file to view it online or download it.
USPI BALL VALVE SAMPLING PROCEDURE
Read more →
USPI SAMPLE PROBE SAMPLING PROCEDURE
Read more →

USPI NH3 PROBE SAMPLING PROCEDURE
Read more →

USPI VISUAL SAMPLING PROCEDURE
Read more →

USPI NH3 VISUAL SAMPLING PROCEDURE
Read more →

IMPROPER OIL SAMPLING PROCEDURES
Read more →
Ball Valve Sampling Procedure
- Only sample running machinery operating at normal/optimum temperatures.
- Never sample machinery that is not running or shortly after machinery start-up.
- Remove 1/4” plug from the sampling valve.
- You may need to wipe clean the sample valve.
- PURGE 4 oz. of oil by slowly moving the ball valve handle in a downward direction. DISCARD AND DO NOT SAMPLE THIS OIL! This cleans the oil sampling valve and piping leading to the sampling valve of any “heavies” or water that may accumulate in low or idle areas of oil flow.
- Due to the extremely small particulate size being measured via oil analysis, please do not remove the oil sample bottle cap until the final sample is ready to be drawn into the bottle. Airborne solids and moisture can enter the sample bottle and skew the lab analysis.
- Fill each sample completely to the top of the bottle, firmly closing lid (note: NH3 samples must first be vented by applying lid loosely before tightening).
- Insert and tighten 1/4” plug back into ball valve.
- Complete the sample label you received from USPI including: sample date, fluid and machinery hours, serial numbers, etc.
- Immediately send completed oil samples to the appropriate oil analysis lab.
Sample Probe Sampling Procedure
- Only sample running machinery operating at normal/optimum temperatures.
- Never sample machinery that is not running or shortly after machinery start-up.
- Remove sample probe cap and loosen protector nut.
- You may need to wipe clean the sample probe.
- PURGE 4 oz. of oil by slowly pushing the sample probe button. DISCARD AND DO NOT SAMPLE THIS OIL! This cleans the oil sampling probe and piping leading to the sample probe of any “heavies” or water that may accumulate in low or idle areas of oil flow.
- Due to the extremely small particulate size being measured via oil analysis, please do not remove the oil sample bottle cap until the final sample is ready to be drawn into the bottle. Airborne solids and moisture can enter the sample bottle and skew the lab analysis.
- Fill each sample completely to the top of the bottle, firmly closing lid (note: NH3 samples must first be vented by applying lid loosely before tightening).
- Replace and tighten sample probe cap and thread forward protector nut (do NOT overtighten!).
- Complete the sample label you received from USPI including: sample date, fluid and machinery hours, serial numbers, etc.
- Immediately send completed oil samples to the appropriate oil analysis lab.
NH3 Sample Probe Sampling Procedure
- Only sample running machinery operating at normal/optimum temperatures.
- Never sample machinery that is not running or shortly after machinery start-up.
- Remove sample probe cap.
- You may need to wipe clean the sample probe.
- Slightly loosen protector nut enough to allow a small stream of oil. Loosening the nut too much will result in a greater oil flow that normally produces a messy foam/oil mixture.
- PURGE 4 oz. of oil by slowly pushing in the sample probe button. DISCARD AND DO NOT SAMPLE THIS OIL! This cleans the oil sampling probe and piping leading to the sample probe of any “heavies” or water that may accumulate in low or idle areas of oil flow.
- Due to the extremely small particulate size being measured via oil analysis , please do not remove the oil sample bottle cap until the final sample is ready to be drawn into the bottle. Airborne solids and moisture can enter the sample bottle and skew the lab analysis.
- Tilt sample probe head to angle the oil stream directly at the inside of the sample bottle. Fill each sample completely to the top of the bottle, firmly closing lid. Strike the bottle on the ground 4-5 times to dissipate foam (this allows quick sampling).
- Take cap off and repeat process 3-4 times or until the sample bottle is full. NH3 samples must first be vented by applying lid loosely before tightening fully.
- Replace and tighten sample probe cap and thread forward protector nut (do NOT overtighten!).
- Complete the sample label you received from USPI including: sample date, fluid and machinery hours, serial numbers, etc.
- Immediately send completed oil samples to WearCheck (address is on sample bottle) for oil analysis.
Improper Oil Sampling Procedures
The greatest hindrance to a successful oil analysis program is improper oil sampling procedures. Failure to follow the recommendations in this chapter will make the oil analysis program cost more time and money and lower its effectiveness. Re-sampling requirements incur labor and financial costs. Re-sampling also delays lab results, which may indicate the need for critical action, and increases the potential for unnecessary equipment wear increases as maintenance lags. Finally, improperly taken samples may indicate the need for filter or fluid changes when none are necessary.
The following common problems may lead to a skewed oil sample:
- Dirty hands, gloves or sampling valve
- Dirty atmosphere when sample was taken.
- Sampling bottle lid left off for extended period of time allowing airborne contaminants and moisture to enter bottle
- Sampling equipment while it is not running
- Sampling before equipment has run long enough to reach operational temperature
- Not purging the sampling valve
- Taking the sample from purged oil